An intensive study of conservators requirements and the many limitations of current dataloggers in the market place has resulted in the development and conception of the rh-t bug. Hanwell have incorporated their wealth of experience in environmental monitoring and the latest components and technology in the creation of a product with a unique combination of hard/software benefits. The result is a specification associated with top of the range instrumentation available in a small and cost effective package. Compare for yourself.
USER CALIBRATION
A major disadvantage with other loggers is the necessity to return the whole unit to the manufacturer for checking and calibration. This costly and time wasting procedure has been eliminated. The rh-t bug has been designed so that the user can check on its performance on a regular basis and calibrate up to and including national standards. Temperature and humidity can be calibrated at up to three points through the software. A humidity reference can be provided by the external connection of industry standard salts, which are simply pushed onto the top of the rh-t bug.
MEMORY
16k of incorruptible EEPROM memory records up to 8184 pairs of humidity and temperature readings. This corresponds to one year's recording time at a logging interval of one hour.
HIGH SPEED COMMUNICATIONS
The rh-t bug provides full specification RS232 comms. at 19200 Baud, while most other products provide only marginal levels at a much lower Baud rate. The impressive result is an entire memory download in less than 15 seconds.
 |
|
Specification
|
 |
(Accuracy quoted is combined instrument and sensor)
|
| Dimensions |
67mm x 67mm x 27mms
|
| Weight |
65 grams
|
| Power supply |
3.6 volt lithium batteries
|
| Battery life |
Up to 5 years (depending upon use)
|
Humidity sensor
|
Capacitive polymer
|
| Temperature sensor |
Precision Thermistor
|
| Logging intervals |
Programmable from 4 seconds to 18 hours
|
| Memory |
16k EEPROM
|
| Record capacity |
16368 records
|
| Delayed start |
Yes
|
|
|
Temperature
|
|
| Range |
-25°C to +80°C
|
| Accuracy |
-25 to -5 +/-0.3°C
-5 to +50 +/-0.2°C
+50 to +70 +/-0.3°C
+70 to +80 +/-0.4°C
|
|
|
Humidity
|
|
| Range |
0 to 100 % non condensing |
| Accuracy |
+/- 2 % over full range |

|
|